Columns
ODS-AL is a non-endcapped high carbon load monomeric C18 phase. It is used only in those situations where residual sianol activity is beneficial to the separation. It is not recommended for the analysis of amines or other compounds with basic functional groups present as peak tailing may occur.
Particle size | : 5µm |
---|---|
Pore size | : 120Å |
Carbon contents | : 17% |
Usable pH range | : 2.0-7.5 |
Column size | : 150X4.6mmI..D. (5µm, 120Å) |
---|---|
Eluent | : 100mM KH2PO4-Na2HPO4(pH 7.0) / methanol(75/25) |
Flow Rate | : 0.8mL/min |
Temperature | : 30ºC |
Detection | : UV at 270nm |