YMC-Triart Series

YMC-Triart series 1.9 µm columns based on novel, multi-layered, organic/inorganic hybrid silica particles are designed for UHPLC. The outstanding mechanical strength of Triart series 1.9 µm makes it ideal for use at higher pressures typical of UHPLC.

Technical data

Features

Suitable for ultra-fast separation

YMC-Triart series 1.9 µm offers high resolution with short length column. Since its efficiency does not decrease even if the flow rate increases, it enables ultra high throughput analysis.

Ultra-fast separation of drugs

  1. Hydrochlorothiazide
  2. Valsartan
  3. Losartan potassium
  4. Amlodipine besilate
  5. Atorvastatin calcium hydrate
  6. Candesartan cilexetil
Eluent A) 10 mM CH3COONH4-CH3COOH (pH 5.5)
B) acetonitrile
Detection UV at 254 nm
Temperature 30℃
Injection 2 µL

Useful for high resolution analysis

YMC-Triart series 1.9 µm has superior column efficiency. It is suitable for multi-component separation such as peptide mapping that requires high resolution.

Peptide mapping

Column YMC-Triart C18
Eluent A) water/TFA (100/0.1)
B) acetonitrile/TFA (100/0.08)
5-40%B (0-15 min) for a single column
5-40%B (0-30 min) for two coupled columns
Flow rate 0.4 mL/min
Temperature 70℃
Detection UV at 220 nm
Injection 10 µL for a single column
20 µL for two coupled columns
Sample Tryptic digest of Bovine Hemoglobin
System Agilent 1290

Excellent resolution without adsorption and tailing

YMC-Triart series 1.9 µm's innovative surface modification technology results in excellent peak shapes even for the basic and coordination compounds that often exhibit adsorption or tailing shapes on other UHPLC columns. Thus, its quantitative capability is sufficient, and decreases the risk of missing unknown constituents. YMC-Triart series stands head and shoulders above any other columns in quantitative capability.

Comparison of chromatographic performance for ionic compounds with UHPLC column

Basic compounds
Column 50 X 2.0 mmI.D. or 50 X 2.1 mmI.D.
Eluent 20 mM KH2PO4-K2HPO4 (pH 6.9)/acetonitrile (65/35)
Flow rate 0.2 mL/min
Temperature 40℃
Detection UV at 235 nm
Sample 1. Chlorpheniramine
2. Dextromethorphan
3. Propyl paraben (I.S.)
Coordination compound
Column 50 X 2.0 mmI.D. or 50 X 2.1 mmI.D.
Eluent acetonitrile/0.1% H3PO4 (40/60)
Flow rate 0.2 mL/min
Temperature 40℃
Detection UV at 254 nm
Sample 1. Hinokitiol
2. Methyl benzoate (I.S.)