Laboratory scale Preparative HPLC System
LC-Forte/R-II
- Specifications
-
Basic specification
|
LC-Forte/R-II |
Flow rate range |
0.1-80 mL/min |
Accuracy of flow rate |
±2% (over 5 mL/min) |
System pressure limit |
30 MPa (0.1-50 mL/min), 20 MPa (50-80 mL/min) |
LCD touch screen (on main body) |
10.4 inch
OS : Windows Embedded Standard 7 or
Windows 10 IoT |
Control |
LCD touch screen or interlinked PC |
Stacking injection |
Suctioned by pump |
Manual injection |
By accompanying injector loop (5 mL as standard) |
Manual/Auto-mode |
Operation for injection, recycle and fraction |
Auto-replay |
Log-expanding auto separation method |
Cleanup |
Auto-flushing for flow line, column-flushing |
Recycle |
By switching of electromagnetic-valve |
PC control (*optional) |
By dedicated software, with LAN connection |
Event marking |
Any events (start, inject, fraction, etc.) |
Timer for pump |
Automatically start/stop at the input time |
Pressure limiter |
Available at any setting values |
Display for chromatogram |
On LCD |
Storage of operation data |
7 GB internal memory/flash device or interlinked PC |
Port |
LAN: 1, USB: 2 |
Sensor for leaking |
Installed |
Storage of detector |
In the enclosure (ALL IN ONE) |
Column holder |
Installed as standard |
Dimension |
270 (W) x 570 (D) x 675 (H) mm |
Power supply |
AC100-240 V (400 W) |
Weight |
Approx. 40 kg |
Detector (selectable)
|
YUV-3400 |
YRI-8830 |
UV |
RI |
Wavelength |
200-400 nm (3 Variable wavelengths) |
- |
Source of light |
D2 lamp |
LED |
Sensitivity/noise |
1.3 x 10-4 AU |
1.4 x 10-6 RIU |
Drift |
3.0 x 10-3 AU/h |
7.9 x 10-6 RIU/h |
Flow cell |
- |
8 µL |
Saturated concentration |
- |
4 x 10-3 RIU以上 |
Measuring method |
Single-beam transmissive-type |
Deflection |
RI range |
- |
1.00-1.75 |
Purge process |
- |
Automatic |
*The standard YUV-3400 is downgradable to other detectors depending on the intended purpose.
Fraction Collector (optional)
|
Forte/FR |
Fraction method |
Replay, manual, peak (controlled by interlinked PC) |
Number of fraction |
Max. 65 fractions (standard rack: 21 fractions) |
Rack size |
φ18 x 65, φ40 x 21 (standard), φ105 x 6, customizable for other size |
Dimension |
300 (W) x 500 (D) x 300 (H) mm |
Power supply |
AC100-240 V |
Weight |
15 kg |
Buffer Rack (optional)
|
Forte/RK |
Column holder |
Installed as standard |
Dimension |
400 (W) x 500 (D) x 420 (H) mm |
Weight |
8 kg |